+44 (0) 1865 400321

Mon-Thurs: 09:00-17:30 Fri: 09:00-16:30

Julabo DYNEO DD-310F 3 to 4 Litre Refrigerated and Heating Circulator, Natural Refrigerant

SKU: 4AJ-4685634
BRAND:

POA

The Julabo DYNEO DD-310F refrigerated and heating circulator is ideal for basic research, materials testing, technical systems and more. This model is suitable for internal and external applications with easy switch over between the two.

Use this quote list to submit your required products to Medline in order for a quotation.

Description

The Julabo DYNEO DD-310F refrigerated and heating circulator offers high heating/cooling capacities for short heat-up and cool-down times. The DD-310F uses environmentally friendly natural refrigerant (R290) and its space-saving cooling coil design delivers more usable space in the bath tank.

Features

Suitable for internal and external applications with simple pump switch over

Powerful and infinitely adjustable pressure pump

Flow rate 27 l/min, pressure 0.7 bar

Large colour TFT display, multilingual

3 Point calibration

Central rotary knob (controller) simplifies operation

Integrated programmer: 8×60 steps

Integrated external Pt100 connection

USB connection

RS232 interface or analogue interfaces (optional)

Bath lid and drain tap included

Removable ventilation grid

Class III (FL) according to DIN 12876-1

High-quality bath tanks made of stainless steel

Specification

Temperature (°C)

Range: -30°C to +200

Setting Resolution: 0.01

Stability: ±0.01

Volume (L): 3 to 4

Pump

Flow Pressure (bar): 0.1 to 0.7

Flow Rate (l/min): 8 to 27

Heating Capacity (kW): 2

Cooling Capacity (Ethanol) (kW)

@ 20°C: 0.30

@ 0°C: 0.27

@ -10°C: 0.21

@ -20°C: 0.12

@ -30°C: 0.02

Timer (hr.min): 0.00 to 99.59

Usable Bath Opening (W x L / D) (mm): 130 x 150 / 150

External Dimensions W x L x H (mm): 230 x 400 x 650

Weight (Kg): 27.4

Barbed Fittings (ID) (mm): 8/12

Pump Connections: M16 x 1 (male)

Ambient Temperature (°C): +5 to +40

Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20°C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Note about natural refrigerants:

Temperature control units using natural refrigerants are often subject to regulatory requirements regarding the installation site, operation, transport or disposal of the units.

Ask a Question